Application of Detector729 in radiotherapy dosimetry
DOI:
CSTR:
Author:
Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective:To study accuracy of 2-D matrix Detector729 in dosimetric measurement and its application in radiotherapy quality control. Methods:Tests and output factor measurement were performed to evaluate and compare the accuracy of Detector729 taking 0.6 cm3 ion chamber under the X-ray as reference. Detector729 was used as a tool to verify dosimetric distribution for inten-sity modulated radiation therapy(IMRT). Results:The scale factor of Detector729 from the reference ionization chamber was 0.991,and the comparison results showed the maximum deviation of measuring point dose was 0.57%. Meanwhile,under the square fields of 4 cm×4 cm-20 cm×20 cm,the estimated output factor matched very well with the reference ion chamber measurements. Conclusion:With high accuracy and reliability,Detector729 can be used in dosimetric measurement and quality control of radiotherapy and pro-vide theoretical basis for the standard use of radiation dosimetric measurement instruments.